Labmate NXT

Next Generation Semi-Automated Biochemistry & Immuno Turbidimetry Analyzer High-performance, semi-automated analyzer designed for accurate, reliable, and efficient biochemistry testing in clinical laboratories.

Overview

Overview

Labmate NXT is a smart choice for laboratories seeking a versatile, semi-automated biochemistry analyzer. Designed for routine chemistries, electrolytes, and drug tests, it combines advanced optics, analytical versatility, and intuitive software to simplify testing and reporting.

  • 8 modes of operation for flexible test management
  • Real-time curve monitoring and patient result tracking
  • Collated reports by Patient ID, Date, or Test
  • Unlimited patient result storage with daily and monthly QC graphs
  • 7-inch capacitive touch color display for smooth navigation
  • Intended for quantitative and qualitative in vitro testing of analytes in body fluids under trained supervision
Features

Key Features

  • High Precision : Maintenance-free photometer ensures accurate and precise results.
  • Direct Reporting : Print reports directly on letterhead through Labmate NXT.
  • Flexible Workload : Ideal for small to medium laboratories.
  • Lamp Saver Mode : Extends lamp life for long-term reliability.
  • Comprehensive Storage : Unlimited patient results and 4-level quality control.

Technical Specifications

Optical Unit

  • 12V, 20W Quartz Halogen Lamp
  • Monochromatic and Bichromatic measurement
  • Absorbance range : 0–3.5 Abs
  • 6 filters (Optional 2 : 450nm & 700nm) — 340nm, 405nm, 505nm, 546nm, 578nm, 630nm
  • Rotating filter wheel mechanism using stepper motor

Measuring Unit

  • 32 μL Quartz Flow Cell
  • Peltier temperature control : 20°C–40°C (steps of 1°C)
  • Digitally controlled using Peltier module and sensor
  • Patient results collated by ID, Date, and Test
  • Aspiration volume: 200–1000 μL

Display & Incubator

  • 7-inch capacitive touch color display
  • Inbuilt incubator with 8 positions for standard round test tubes
  • Fixed at 37°C ±0.5°C, digitally controlled using silicone heater pad and sensor

Analysis Methods

  • End Point, Fixed Time, Fixed Time with Multi-Standards
  • Kinetic, Kinetic with Standard, Multi Standard Absorbance

Benefits

  • Accurate and reliable results every time
  • Easy report generation and data management
  • Flexible for varying laboratory workloads
  • Long-term operational reliability with lamp saver and maintenance-free design
  • Complete QC monitoring for consistent test quality

Need Guidance?
Our Experts Are Here for You.